Sign In | Join Free | My udpf.com
China Labnovation Technologies, Inc. logo
Labnovation Technologies, Inc.
Labnovation Technologies,Inc.
Active Member

4 Years

Home > HPLC HbA1c Analyzer >

AC 100 - 240V Automated Hba1c Analyzer High Performance HPLC Method For Diabetes Diagnosis

Labnovation Technologies, Inc.
Contact Now

AC 100 - 240V Automated Hba1c Analyzer High Performance HPLC Method For Diabetes Diagnosis

Brand Name : LABNOVATION

Model Number : LD-560

Certification : IFCC, NGSP, CE

Place of Origin : Shenzhen, China

MOQ : 1 Unit

Price : USD7,000.00-12,000.00/Unit

Supply Ability : 100 Units/Day

Delivery Time : 15-20 working days

Packaging Details : Carton Size: 480mm×480mm×615mm, Gross Weight: 28kgs

Principle : HPLC (High Performance Liquid Chromatography)

Test Speed : 90s/test

CV : <2%

Sample types : Whole blood or Pre-diluent sample

Test blood volume : 5μL in whole blood mode; 150μL in pre-diluted mode

Sample loading method : Continuous automatic sample loading

Sample capacity : 20 samples

Reagent pack : 600T

Column test quantity : 3000T

Eluent Kind : Eluent A, Eluent B,Eluent C (Variant Mode)

Data storage : 5000

Printer : Built-in Thermal Printer

Test range : 3.0%~18.0%

Operation screen : 10" Full color LCD display

Gross Weiht : 28kg

Dimensions : 378(L) ×380(W)×510(H)mm

Shelf Life : 5 Years

Warranty of Device : 1 Year

Warranty of Sampling Sensor : 6 Month

Contact Now

Automatic HbA1c Analyzer High Performance HPLC Golden Method for Diabetes Diagnosis


Intended use


The Fully Automated HbA1c Analyzer is used in conjunction with suitable reagents for the quantitative detection of glycosylated hemoglobin (HbA1c) in human blood.The Fully automatedHbA1c Analyzer LD-560 offers the fast throughput of HbAlc results in 90 seconds, with Hb variant detection, providing the outstanding solution for quick and reliable diabetic monitoring. No sample preparation and very little hands-on time by the operator is required for the LD-560 Analyzer.

Specification


Model LD-560
Methodology High-Performance Liquid Chromatography (HPLC)
Test Modes Fast Mode, Variant Mode
Test Range 3% - 18%
Precision (CV) <2%
Parameter HbA1c, HbF, HbA1a, HbA1b, HbA0, HbA2(Optional)
First Sample Result 4 Mins
Test Speed Fast Mode: 90 Seconds/Test
Variant Mode: 140 Seconds/Test
Sample Type Whole blood, Pre-diluted blood
Sample Volume Whole blood: 5μL
Pre-diluted blood: 150μL
Sample loading capacity 20
Photometer 420nm
Chromatography Column Capacity 3000T
Filter Capacity 600T
Display 10“ LCD touch screen
Reagent Kit Eluent A, Eluent B, Eluent C, Hemolysis, Chromatography Column
Storage 5000 Sample Results
Connection interface USB, LAN, LIS Compatible
Printer Thermal Printer
System operation environment Temperature 15-30℃
Humidity 30%-85%
Power AC 100-240V 50/60HZ 300VA
Dimensions 378mmx380mmx510mm
Weight 28kgs

HPLC Methodology
High-Pressureliquid Chromatography (HPLC), to separate HbAlc directly with measuring the absor bance points continually to form chromatogram. Using normal distribution curve fitting auto-iterative algorithm to get precise HbAlc testing result, excluding interference of variant and unstable hemoglobin like HbF. Standard Analysis Mode will report HbAla,HbAlb,HbF,HbAlc,LAlc,HbAO peak areas and ratio. And the result also includes IFCC, NGSP and ADAG value for diverse client needs.
Performance

*HPLC, CV<2% - Accurate and Stable

*Full Automatic smart System
*Quality & Performance - IFCC & NGSP Certificate
*User Friendly

  • 10-30 daily samples
  • 90 seconds / Test
  • 10” LCD touch screen
  • 28 kgs only

Product Composition

Main Parts

The analyzer consists of an auto sample injection module, a liquid chromatography separation module, a colorimetric detection and control module, a barcode scanning module, and a display printing module.



Reagents and Consumables

The analyzer is recommended to be used in conjunction with the Glycosylated Hemoglobin Test Reagent Kit (high performance liquid chromatography (HPLC method)) produced by our company, which includes the following components:
Eluent A/Eluent B/Eluent C(Variant Mode) Adjusts the pH of the solution for the elution of glycosylated hemoglobin and non glycated hemoglobin.
Hemolysis Reagent Used to dissolve red blood cells and release hemoglobin.
Chromatography Column Used for separation and elution of glycosylated hemoglobin and non glycated.
Hemoglobin Filter Used to initially filter the sample after hemolysis.
Paper Rolls Prints a report of the test results.
AC 100 - 240V Automated Hba1c Analyzer High Performance HPLC Method For Diabetes Diagnosis

Certificate


AC 100 - 240V Automated Hba1c Analyzer High Performance HPLC Method For Diabetes Diagnosis


Product Tags:

AC240V HbA1c HPLC analyzer

      

Thermal Printer HbA1c HPLC analyzer

      

HPLC Method automated hba1c analyzer

      
Best AC 100 - 240V Automated Hba1c Analyzer High Performance HPLC Method For Diabetes Diagnosis wholesale

AC 100 - 240V Automated Hba1c Analyzer High Performance HPLC Method For Diabetes Diagnosis Images

Inquiry Cart 0
Send your message to this supplier
 
*From:
*To: Labnovation Technologies, Inc.
*Subject:
*Message:
Characters Remaining: (0/3000)